Development of an Automated Method for Analysis of Winston-Lutz Test Results Using Digital Radiography and Photostimulable Storage Phosphor
Autor: | Kazuya Hayashida, Sunao Ohkura, Kazutaka Nashiki, Kousuke Nagafuchi, Tomomi Kawahara, A. Mizoguchi, Yoshifumi Saida, Ayumu Ohishi, Hidemichi Kawata |
---|---|
Rok vydání: | 2013 |
Předmět: |
Reproducibility
Pixel Brain Neoplasms business.industry Computer science medicine.medical_treatment Template matching Isocenter General Medicine Radiation Radiosurgery Linear particle accelerator Radiographic Image Enhancement Optics Computer graphics (images) Image Processing Computer-Assisted medicine Humans Particle Accelerators business Algorithms Radiotherapy Image-Guided Digital radiography |
Zdroj: | Japanese Journal of Radiological Technology. 69:1266-1273 |
ISSN: | 1881-4883 0369-4305 |
Popis: | Stereotactic radiosurgery (SRS) and radiotherapy (SRT) are intricate techniques that deliver a highly precise radiation dose to a localized target, usually a tumor. At our hospital, we perform SRS and SRT on brain tumors using a linear accelerator (linac) mounted with an external micro multi-leaf system. The Task Group TG-142 Report by the American Association of Physicists in Medicine recommends the coincidence of the radiation and mechanical isocenter to be within ±1 mm. The Winston-Lutz test is commonly used to verify the linac isocenter position: it has the advantages of being a simple method that uses a film or electronic portal imaging device (EPID). However, the film method requires a higher radiation dose, which makes it more time-consuming than the EPID method, and the results are highly dependent on the skills of the observer. The EPID method has certain advantages over the film method, but it has low resolution and can only be used for a few combinations of gantry and couch angles. This prompted us to develop an in-house-designed radiation receptor system based on digital radiography, using a photostimulable storage phosphor and automated analysis algorithm for Winston-Lutz test images using a template-matching technique based on cross-correlation coefficients. Our proposed method shows a maximum average absolute error of 0.222 mm (less than 2 pixels) for 0.5 mm and 1.0 mm displacement from the isocenter toward the inline and crossline directions. Our proposed method is thus potentially useful for verifying the Linac isocenter position with a small error and good reproducibility, as demonstrated by improved accuracy of evaluation. |
Databáze: | OpenAIRE |
Externí odkaz: |