Validation of the BIOS BD240 home blood pressure monitor in a multi-racial, multi-morbid Canadian population according to the International Standards Organization 2018 Protocol

Autor: Mira Wirzba, Fabiola Diaz-Carvallo, Jennifer S Ringrose, Peter W. Wood, Raj Padwal
Rok vydání: 2020
Předmět:
Zdroj: Blood pressure monitoring. 25(3)
ISSN: 1473-5725
Popis: OBJECTIVE Assess the accuracy of the BIOS BD240 home blood pressure (BP) monitor and wide-range cuff according to the International Organization for Standardization (ISO) 2018 standard. METHODS Eighty-five subjects (aged ≥18 y) with arm circumferences between 24 and 43 cm were studied. Blinded, two-observer, mercury-based auscultation performed using a two-piece cuff chosen for upper arm size was used for reference measurements. Accuracy criteria 1 and 2 of the ISO standard were determined and Bland-Altman plots generated. RESULTS Sixty percent of the study sample was female and 42% had hypertension. Mean device-to-reference standard differences in SBP/DBP were 0.0 ± 6.2/-0.2 ± 6.3 for criterion 1 and 0.0 ± 4.8/-0.2 ± 5.8 for criterion 2. CONCLUSION The BIOS BD240 passed the requirements of the American National Standards Institute/Association for the Advancement of Medical Instrumentation/ISO standard for both SBP and DBP and can be recommended for use.