Survival Rate of Cells Sent by a Low Mechanical Load Tube Pump: The 'Ring Pump'

Autor: Inoue Hiroaki, Kaoru Uesugi, Yoji Yamada, Yoichi Sakurai, Keisuke Morishima, Koki Hirai, Keizo Nishiyama, Takashi Taniguchi
Jazyk: angličtina
Rok vydání: 2020
Předmět:
Zdroj: Micromachines
Volume 11
Issue 4
Micromachines, Vol 11, Iss 447, p 447 (2020)
ISSN: 2072-666X
DOI: 10.3390/mi11040447
Popis: A ring pump (RP) is a useful tool for microchannels and automated cell culturing. We have been developing RPs (a full-press ring pump, FRP
and a mid-press ring pump, MRP). However, damage to cells which were sent by the RP and the MRP was not investigated, and no other studies have compared the damage to cells between RPs and peristaltic pumps (PPs). Therefore, first, we evaluated the damage to cells that were sent by a small size FRP (s-FRP) and small size MRPs (s-MRPs
gap = 25 or 50 &mu
m, respectively). &ldquo
Small size&rdquo
means that the s-FRP and the s-MRPs are suitable for microchannel-scale applications. The survival rate of cells sent by the s-MRPs was higher than those sent by the s-FRP, and less damage caused by the former. Second, we compared the survival rate of cells that were sent by a large size FRP (l-FRP), a large size MRP (l-MRP) (gap = 50 &mu
m) and a PP. &ldquo
Large size&rdquo
means that the l-FRP and the l-MRP are suitable for automated cell culture system applications. We could not confirm any differences among the cell survival rates. On the other hand, when cells suspended in Dulbecco&rsquo
s phosphate-buffered saline solution were circulated with the l-MRP (gap = 50 &mu
m) and the PP, we confirmed a difference in cell survival rate, and less damage caused by the former.