A high-frame-rate ultrasound system for the study of tissue motions
Autor: | Robert Rohling, Septimiu E. Salcudean, Alex D. Brant, Ali Baghani |
---|---|
Rok vydání: | 2010 |
Předmět: |
Acoustics and Ultrasonics
Computer science Movement Acoustics ComputingMethodologies_IMAGEPROCESSINGANDCOMPUTERVISION Models Biological symbols.namesake Optics Motion estimation Image Processing Computer-Assisted medicine Humans Ultrasonography Doppler Color Electrical and Electronic Engineering Instrumentation ComputingMethodologies_COMPUTERGRAPHICS Motion compensation medicine.diagnostic_test Phantoms Imaging Color image business.industry Ultrasound Frame rate Time–frequency analysis symbols Elastography business Doppler effect Algorithms |
Zdroj: | IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control. 57:1535-1547 |
ISSN: | 0885-3010 |
DOI: | 10.1109/tuffc.2010.1584 |
Popis: | In this article, a technique for measuring fast periodic motion is proposed. The sequencing used in this technique is similar to the one used in conventional color Doppler systems. However, a phase correction algorithm is introduced which compensates for the acquisition delays. Criteria for the types of motion which could be detected correctly by the system are developed and presented. Effective frame rates of several hundred hertz to a few kilohertz have been achieved with the system. Applications of the system in tissue elastography are presented together with experimental results from tissue mimicking phantoms. |
Databáze: | OpenAIRE |
Externí odkaz: |