Evaluation of manual resuscitators used in ICUs in Brazil*
Autor: | Carlos Roberto Ribeiro de Carvalho, Márcia Souza Volpe, Tatiana de Arruda Ortiz, Mauro R. Tucci, Marcelo B. P. Amato, Germano Forti Junior, Marcelo A. Beraldo |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2013 |
Předmět: |
Pulmonary and Respiratory Medicine
Segurança de equipamentos Male medicine.medical_specialty Comparative Effectiveness Research Resuscitation Respiracao artificial Materials testing Seguranca de equipamentos Desenho de equipamento medicine Tidal Volume Humans Computer Simulation Respiracao com pressao positiva Positive-Pressure Respiration Tidal volume Teste de materiais lcsh:RC705-779 Observer Variation Positive-pressure respiration Manual resuscitators Ventilators Mechanical business.industry Equipment safety Artigo Original lcsh:Diseases of the respiratory system Original Articles Surgery Intensive Care Units Respiração artificial Anesthesia Safety Equipment Equipment design Respiratory Mechanics Female Ressuscitação Respiration artificial business Respiração com pressão positiva Ressuscitacao Brazil |
Zdroj: | Jornal Brasileiro de Pneumologia : Publicaça̋o Oficial da Sociedade Brasileira de Pneumologia e Tisilogia Jornal Brasileiro de Pneumologia v.39 n.5 2013 Jornal Brasileiro de Pneumologia Sociedade Brasileira de Pneumologia e Tisiologia (SBPT) instacron:SBPT Jornal Brasileiro de Pneumologia, Vol 39, Iss 5, Pp 595-603 (2013) Jornal Brasileiro de Pneumologia, Volume: 39, Issue: 5, Pages: 595-603, Published: SEP 2013 |
ISSN: | 1806-3756 1806-3713 |
Popis: | OBJECTIVE: To evaluate the performance of manual resuscitators (MRs) used in Brazil in accordance with international standards. METHODS: Using a respiratory system simulator, four volunteer physiotherapists employed eight MRs (five produced in Brazil and three produced abroad), which were tested for inspiratory and expiratory resistance of the patient valve; functioning of the pressure-limiting valve; and tidal volume (VT) generated when the one-handed and two-handed techniques were used. The tests were performed and analyzed in accordance with the American Society for Testing and Materials (ASTM) F920-93 criteria. RESULTS: Expiratory resistance was greater than 6 cmH2O . L−1 . s−1 in only one MR. The pressure-limiting valve, a feature of five of the MRs, opened at low pressures (< 17 cmH2O), and the maximal pressure was 32.0-55.9 cmH2O. Mean VT varied greatly among the MRs tested. The mean VT values generated with the one-handed technique were lower than the 600 mL recommended by the ASTM. In the situations studied, mean VT was generally lower from the Brazilian-made MRs that had a pressure-limiting valve. CONCLUSIONS: The resistances imposed by the patient valve met the ASTM criteria in all but one of the MRs tested. The pressure-limiting valves of the Brazilian-made MRs usually opened at low pressures, providing lower VT values in the situations studied, especially when the one-handed technique was used, suggesting that both hands should be used and that the pressure-limiting valve should be closed whenever possible. OBJETIVO: Avaliar o desempenho de reanimadores manuais (RMs) utilizados no Brasil conforme critérios definidos por uma norma internacional. MÉTODOS: Utilizando um simulador do sistema respiratório, oito RMs manuais (cinco produzidos no Brasil e três importados) foram avaliados em relação a resistência inspiratória e expiratória da válvula para o paciente; funcionamento da válvula limitadora de pressão; e volume corrente (VT) gerado por quatro fisioterapeutas voluntários, utilizando uma ou duas mãos. Para a realização e análise dos testes, foram utilizados critérios sugeridos pela norma da American Society for Testing and Materials (ASTM) F920-93. RESULTADOS: A resistência expiratória foi superior a 6 cmH2O . L−1 . s−1 em apenas um dos RMs testados. A válvula limitadora de pressão, presente em cinco RMs, abriu em baixas pressões (< 17 cmH2O), e a pressão máxima variou de 32,0-55,9 cmH2O. Houve grande variação da média de VT obtido pelos diferentes RMs. Os valores médios de VT utilizando uma mão foram inferiores ao sugerido pela ASTM (600 mL). O VT médio, nas situações estudadas, foi geralmente menor nos RMs nacionais com válvula limitadora de pressão. CONCLUSÕES: As resistências impostas pela válvula do paciente estão de acordo com os critérios da ASTM, com exceção de um RM. As válvulas limitadoras de pressão dos RMs nacionais geralmente abrem em baixas pressões, determinando o fornecimento de um menor VT nas situações estudadas, principalmente com o uso de uma mão, o que sugere que a ventilação deva ser feita com as duas mãos e a válvula limitadora de pressão deva ser fechada sempre que possível. |
Databáze: | OpenAIRE |
Externí odkaz: |