Validation of a Lateral Flow Assay for Rapid Diagnosis of Histoplasmosis in Advanced HIV Disease, Buenos Aires, Argentina

Autor: Mariana Andreani, Claudia E. Frola, Diego H. Caceres, Cristina E. Canteros, María J. Rolón, Tom Chiller, Liliana Guelfand
Jazyk: angličtina
Rok vydání: 2022
Předmět:
Zdroj: Applied Microbiology; Volume 2; Issue 4; Pages: 950-955
ISSN: 2673-8007
DOI: 10.3390/applmicrobiol2040072
Popis: Histoplasmosis is a major cause of mortality in individuals with advanced human immunodeficiency virus (HIV) disease (AHD). We evaluated in patients with AHD a lateral flow assay (LFA) developed by MiraVista® Diagnostics (MVD LFA). Histoplasmosis was defined based on the European Organization for Research and Treatment of Cancer/Invasive Fungal Infections Cooperative Group and the National Institute of Allergy and Infectious Diseases Mycoses Study Group (EORTC/MSG) case definitions. We also compared the results of this LFA with those obtained using a commercial enzyme immunoassay (EIA) developed by IMMY, Clarus Histoplasma GM EIA, IMMY (HGM EIA). A retrospective observational study was conducted at Hospital Juan A. Fernández, located in Buenos Aires, Argentina. The study included 48 urine specimens from patients aged >18 years with AHD. Urine specimens included 17 patients with disseminated histoplasmosis and 31 specimens from patients without evidence of histoplasmosis. Specimens were tested using the MVD LFA and the HGM EIA. The MVD LFA and the HGM EIA had similar analytical performance, with a sensitivity of 94%, specificity of 100%, positive predictive value of 100%, negative predictive value of 97%, and an accuracy of 98%. Comparison of the MVD LFA with the HGM EIA demonstrated a Kappa agreement index of 0.906. The LFA evaluated in this study had high analytical performance; it provided rapid diagnosis of histoplasmosis with minimal requirements for laboratory training, equipment, and laboratory infrastructure.
Databáze: OpenAIRE
Nepřihlášeným uživatelům se plný text nezobrazuje