Multicenter Evaluation of a Modular Dual Mobility Construct for Revision Total Hip Arthroplasty

Autor: William J. Hozack, Ronald C. Huang, Geoffrey H. Westrich, Steven F. Harwin, Michael A. Mont, Carlos A. Higuera-Rueda, Art L. Malkani
Rok vydání: 2018
Předmět:
Zdroj: The Journal of arthroplasty. 34(7S)
ISSN: 1532-8406
Popis: Introduction Modular Dual Mobility (MDM) constructs in total hip arthroplasty (THA) offer increased hip stability compared with constrained liners, without compromising hip range of motion. The purpose of this study was to evaluate outcomes of revision THA using MDM. Methods The study was a multiinstitutional retrospective cohort study of 315 hips that underwent revision THA using MDM between 2011 and 2017. Clinical outcomes and reasons for failure were collected. Results Three hundred fifteen patients met 1-year minimum follow-up (mean 3.3 years). Nine hips had instability postoperatively (2.9%), and 30 hips required reoperation (9.5%). Seven had recurrent instability (6.5%) which was associated with liner-only exchange (P = .021) and liner outer diameter of ≤ 38 mm (P = .016). Conclusion Revision THA with MDM provided a low rate of instability and reoperation in a revision cohort. Recurrent instability following use of MDM in revision THA was associated with retention of the acetabular component and polyethylene outer diameter ≤ 38 mm. Level of Evidence Therapeutic Level III.