Alternative method of intraocular lens power calculation in eyes with short axial length

Autor: D.F. Belov, V.P. Nikolaenko
Rok vydání: 2022
Předmět:
Zdroj: Vestnik oftal'mologii. 138:24
ISSN: 0042-465X
DOI: 10.17116/oftalma202213803124
Popis: To develop an alternative method of intraocular lens (IOL) power calculation in eyes with mature cataract and axial length (AL) of less than 22.0 mm using modern formulas Barrett Universal II and Hill RBF.The study enrolled 41 patients (41 eyes) who underwent phacoemulsification (PE). Ultrasound biometry (Tomey Biometer Al-100) and keratometry (Topcon-8800) were used for IOL power calculation by SRK/T and Haigis formulas. To calculate IOL power by Barrett Universal II and Hill RBF formulas, 0.2 mm were added to AL measured with ultrasonography (retinal thickness). One month after PE, spherical equivalent of refraction was compared with target refraction (calculated by the formulas listed above), and based on that a conclusion was made on the accuracy of calculations.Haigis formula was found to be the least accurate (IOL calculation error -0.39±0.79 D). The calculation error in SRK/T (0.04±0.79 D), Barrett Universal II (0.02±0.79 D) and Hill RBF (-0.05±0.73 D) formulas was much lower. However, among them Hill RBF had the lowest spread of the mean absolute IOL calculation error. Pairwise comparison revealed significant difference of mean IOL calculation error by Haigis formula versus the others. There was no significant difference in the following pairs: SRK/T - Barrett Universal II (The proposed alternative method of IOL power calculation in eyes with mature cataract and short AL using modern formulas (Barrett Universal II and Hill RBF) shows higher accuracy compared to the formulas embedded in ultrasound biometer (SRK/T and Haigis), and can be recommended for use in everyday practice.Разработка алгоритма расчета оптической силы интраокулярных линз (ИОЛ) в глазах со зрелой катарактой и длиной переднезадней оси (ПЗО)22,0 мм с использованием калькуляторов Barrett Universal II и Hill RBF.Исследуемую группу составили пациенты (Наибольшей ошибкой расчета ИОЛ сопровождалось применение формулы Haigis (–0,39±0,79 дптр). Величина отклонения от целевой рефракции при использовании SRK/T (0,04±0,79 дптр), Barrett Universal II (0,02±0,79 дптр) и Hill RBF (–0,05±0,73 дптр) оказалась намного меньшей. Минимальный разброс абсолютной ошибки расчета ИОЛ характерен для уравнения Hill RBF. При попарном сравнении средних значений ошибки расчета формула Haigis существенно отличалась от всех остальных. В парах SRK/T и Barrett Universal II (Предложенный нами альтернативный метод расчета ИОЛ, использующий современные калькуляторы (Barrett Universal II, Hill RBF), показал более высокую точность по сравнению со встроенными в ультразвуковой биометр формулами (SRK/T, Haigis) и может быть использован в повседневной клинической практике для расчета ИОЛ.
Databáze: OpenAIRE