Accuracy and User Performance of a New Blood Glucose Monitoring System

Autor: Leslie J. Klaff, Pragathi Shelat, Phil Vernes, James M. Richardson, Amy Wayland-Smith, Diana Zondorak
Rok vydání: 2020
Předmět:
Zdroj: J Diabetes Sci Technol
ISSN: 1932-2968
Popis: Introduction: Self-monitoring of blood glucose (BG) is important in diabetes management, allowing people with diabetes (PWD) to assess responses to diabetes therapy and to inform if they are attaining their glycemic targets. This study assessed the accuracy and user performance (UP) of a new blood glucose monitoring system (BGMS), CONTOUR®PLUS ELITE, according to International Organization for Standardization (ISO) 15197:2013 criteria and also more stringent criteria. Methods: In laboratory Study 1, capillary fingertip blood samples from 100 PWD were evaluated using the new BGMS. In clinical Study 2, 130 PWD had Yellow Springs Instrument (YSI) analyzer reference measurements against subject-obtained fingertip and palm blood, and trial staff-obtained venous blood. The new BGMS was tested with test strips from three different lots. A UP questionnaire assessed ease of use. Results: Study 1: 100% of combined accuracy results fulfilled ISO criteria (±15 mg/dL at BG 98% of subject- and 100% of trial staff-obtained performance results met ISO criteria. Most subjects (>96%) found the BGMS easy to use. Conclusion: The new BGMS exceeded minimum ISO 15197:2013-specified standards for both accuracy and UP criteria, along with the more stringent accuracy criteria. These data show that this new BGMS can be a useful tool in managing glycemic control for PWD.