Unusual burns of the lower extremities caused by a closed conducting loop in a patient at MR imaging

Autor: G. van Kaick, Jürgen Debus, H. J. Zabel, Marco Essig, Michael V. Knopp
Rok vydání: 1996
Předmět:
Zdroj: Radiology. 200:572-575
ISSN: 1527-1315
0033-8419
Popis: An extremely rare occurrence of third-degree burns was induced in the medial calves of a male patient with unusual anatomy (after resection and radiation therapy of a liposarcoma) during conventional magnetic resonance (MR) imaging on a clinical 1.5-T MR system that operated without any external conductor present and within safe limits. A closed conducting loop was inadvertently created, which caused focal increased temperature at the junction of his calves.
Databáze: OpenAIRE