Air kerma strength characterization of a GZP6 Cobalt-60 brachytherapy source
Autor: | Yasha Makhdoumi, Ali S. Meigooni, Mohammad Taghi Bahreyni Toossi, Mahdi Ghorbani, Mojtaba Taheri, Mohsen Layegh, Ali Asghar Mowlavi |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2010 |
Předmět: |
Task group
medicine.medical_specialty Cancer Research business.industry Cobalt 60 brachytherapy medicine.medical_treatment Brachytherapy GZP6 HDR system MCNP 4C Monte Carlo code Brachytherapy source Article Air kerma strength Kerma Oncology Radiology Nuclear Medicine and imaging medicine Radiology Nuclear Medicine and imaging Medical physics Dose rate Nuclear medicine business In air measurement |
Popis: | BackgroundTask group number 40 (TG-40) of the American Association of Physicists in Medicine (AAPM) has recommended calibration of any brachytherapy source before its clinical use. GZP6 afterloading brachytherapy unit is a 60Co high dose rate (HDR) system recently being used in some of the Iranian radiotherapy centers.AimIn this study air kerma strength (AKS) of 60Co source number three of this unit was estimated by Monte Carlo simulation and in air measurements.Materials and methodsSimulation was performed by employing the MCNP-4C Monte Carlo code. Self-absorption of the source core and its capsule were taken into account when calculating air kerma strength. In-air measurements were performed according to the multiple distance method; where a specially designed jig and a 0.6cm3 Farmer type ionization chamber were used for the measurements. Monte Carlo simulation, in air measurement and GZP6 treatment planning results were compared for primary air kerma strength (as for November 8th 2005).ResultsMonte Carlo calculated and in air measured air kerma strength were respectively equal to 17240.01μGym2h−1 and 16991.83μGym2h−1. The value provided by the GZP6 treatment planning system (TPS) was “15355μGym2h−1”.ConclusionThe calculated and measured AKS values are in good agreement. Calculated-TPS and measured-TPS AKS values are also in agreement within the uncertainties related to our calculation, measurements and those certified by the GZP6 manufacturer. Considering the uncertainties, the TPS value for AKS is validated by our calculations and measurements, however, it is incorporated with a large uncertainty. |
Databáze: | OpenAIRE |
Externí odkaz: |