Evaluation of Two Methods of Rapid Blood-Glucose Monitoring by Unskilled Personnel during Surgery

Autor: B Adelhøj, Sten Madsbad, Dennis Richard Bigler, Jannik Hilsted
Rok vydání: 1984
Předmět:
Zdroj: Acta Anaesthesiologica Scandinavica. 28:649-651
ISSN: 1399-6576
0001-5172
DOI: 10.1111/j.1399-6576.1984.tb02138.x
Popis: The accuracy of two rapid methods of blood-glucose monitoring without (Haemo-glucotest 1-44) and with a reflectance meter (Hypocount B) was compared using a laboratory method. The assessment was carried out by personnel with no previous experience in measuring blood glucose. Eighty-five percent of the 92 measurements obtained with the hypocount B were within +/- 20% of the laboratory glucose values. Using haemo-glucotest 1-44 strips, 74% of the readings were within +/- 20% of the reference laboratory values. For values below 5.5 mmol/l, there was a tendency for results to be too low, with 77% of the readings below laboratory values -20%. All situations with severe hypoglycaemia were detected with both strips. The study also demonstrates the ineffectiveness of s.c. insulin regimens during surgery. Only 47% of the measured blood glucose values were within the range of 5.5-10 mmol/l and two of ten patients had hypoglycaemia with values below 2.5 mmol/l. Therefore glucose monitoring during surgery in diabetics seems justified. This monitoring can be performed adequately with strips.