Autor: |
Emilie Alchourron, Josée Dubois, Guy Cloutier, Nina Stein, Ziad Farhat, Marie-Hélène Roy-Cardinal, Jean-Baptiste Moretti, Chantale Lapierre, Ramy El Jalbout |
Rok vydání: |
2022 |
Předmět: |
|
Zdroj: |
Canadian Association of Radiologists journal = Journal l'Association canadienne des radiologistes. |
ISSN: |
1488-2361 |
Popis: |
Background: Childhood obesity is linked to higher adult mortality and morbidity from atherosclerosis. It is primordial to detect at-risk children earlier-on to prevent disease progression. Carotid intima-media thickness (IMT) is a subclinical radiological marker for early atherosclerosis. B-mode ultrasound is a known technique to assess IMT, but no gold standard technique exists in children. Non-invasive vascular elastography (NIVE) using speckle statistics is an innovative alternative to evaluate IMT and adds by providing translation, strain and shear strain measurements. Validation studies for both techniques lack in children. Purpose: Validate the reproducibility of the 2 techniques in Canadian children. Methods: We conducted a prospective study where anthropometry, blood pressure, IMT and elastography were measured. Six operators obtained 2 measurements for both carotid arteries using both techniques, for a total of 720 measurements. Inter- and intra-class correlation coefficients (ICC) were calculated for each measurement technique and elastography parameters. Results: 30 participants (13.0 ± 1.26 years, 17 girls) were recruited. Twelve were overweight. No significant difference was found in mean IMT between weight groups for either technique ( P = .15 and P = .60). We found excellent inter- (ICC = .98 [95% Confidence Interval (CI): .97; .99]) and intra- (ICC = .90-.93) operator reliability for the B-mode technique, and good inter (ICC = .70 [95% CI: .47; .85]) and intra- (ICC = .71-.91) operator reliability for the NIVE-based technique. Poor reliability was found between techniques (ICC = .30 [95% CI: −.31; .65). For elastography parameters, translation was the most reliable (ICC = .94-.95). Conclusion: IMT measurement is reproducible in children but not between techniques. NIVE gives the advantage of evaluating elastography. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|