Concordance between automated oscillometric measurement of ankle–brachial index and traditional measurement by eco-Doppler in patients without peripheral artery disease
Autor: | Julio Antonio Carbayo-Herencia, Manuel Gómez-Serranillos, Alejandro Villarín-Castro, Luis Miguel Artigao-Ródenas, Carlos Escobar-Cervantes, G.C. Rodríguez-Roca, Antonio Segura-Fragoso, Julio Hernández-Moreno, Javier Carrasco-Flores, Francisco J. Alonso-Moreno |
---|---|
Rok vydání: | 2014 |
Předmět: |
Adult
Male medicine.medical_specialty Intraclass correlation Concordance Population Blood Pressure Disease Peripheral Arterial Disease symbols.namesake Oscillometry Internal medicine Diabetes mellitus Diabetes Mellitus Internal Medicine medicine Humans Ankle Brachial Index Obesity education Aged Dyslipidemias education.field_of_study Primary Health Care business.industry Ultrasonography Doppler General Medicine Middle Aged medicine.disease Pearson product-moment correlation coefficient body regions medicine.anatomical_structure Spain Hypertension symbols Physical therapy Cardiology Female Ankle Cardiology and Cardiovascular Medicine business Dyslipidemia |
Zdroj: | Blood Pressure. 23:270-275 |
ISSN: | 1651-1999 0803-7051 |
DOI: | 10.3109/08037051.2013.876796 |
Popis: | To evaluate the concordance between automated oscillometric measurement (WatchBP® Office ABI) of the ankle- brachial index (ABI) and the traditional measurement by eco-Doppler in a Spanish population without peripheral artery disease attended in primary care.The ABI was determined by both methods in a general population aged ≥ 18 years, from the RICARTO study. The intraclass correlation coefficient was calculated to assess the concordance between both techniques and the Bland-Altman plot was determined to analyze the agreement between them.A total of 322 subjects (mean age 47.7 ± 16.0 years; 54.3% women) were included in the study. With regard to cardiovascular risk factors, 70.5% of subjects had dyslipidemia, 26.7% hypertension, 24.8% obesity, 8.4% diabetes and 25.5% were smokers. Mean ABI measured by eco-Doppler and the automated method were 1.17 ± 0.1 and 1.2 ± 0.1, respectively (mean differences - 0.03 ± 0.09; p0.001). The Pearson correlation coefficient and the intraclass correlation coefficient were in both cases 0.70.The automated oscillometric measurement of ABI is a reliable and useful alternative to conventional eco-Doppler determination in the general population without peripheral artery disease attended in primary care. |
Databáze: | OpenAIRE |
Externí odkaz: | |
Nepřihlášeným uživatelům se plný text nezobrazuje | K zobrazení výsledku je třeba se přihlásit. |