Transient out‐of‐range impedance in 'hybrid' implantable cardioverter‐defibrilator system: A case series
Autor: | Wojciech Kwaśniewski, Michał Orszulak, Artur Filipecki, Katarzyna Mizia-Stec, Dagmara Urbańczyk-Swić, Witold Orszulak |
---|---|
Rok vydání: | 2021 |
Předmět: |
Male
medicine.medical_specialty business.industry General Medicine Prosthesis Design System a Defibrillators Implantable Diagnosis Differential Equipment Failure Analysis Internal medicine Electric Impedance medicine Cardiology Retrospective analysis Humans Female Transient (oscillation) Cardiology and Cardiovascular Medicine business Aged Retrospective Studies |
Zdroj: | Pacing and Clinical Electrophysiology. 45:270-273 |
ISSN: | 1540-8159 0147-8389 |
DOI: | 10.1111/pace.14401 |
Popis: | A retrospective analysis of 60 patients with hybrid ICD systems: Boston Scientific device paired with non-Boston leads. In 10 (17%) patients transient, out-of-range peaks of ventricular pace impedance trend were observed. Probable cause is header-lead interaction incompatibility. This matter is known mainly for pacemakers systems but not for ICDs. Investigation this issue is crucial because consequences in ICD systems are unpredictable and risk might be higher than in pacing systems. This article is protected by copyright. All rights reserved. |