Transient out‐of‐range impedance in 'hybrid' implantable cardioverter‐defibrilator system: A case series

Autor: Wojciech Kwaśniewski, Michał Orszulak, Artur Filipecki, Katarzyna Mizia-Stec, Dagmara Urbańczyk-Swić, Witold Orszulak
Rok vydání: 2021
Předmět:
Zdroj: Pacing and Clinical Electrophysiology. 45:270-273
ISSN: 1540-8159
0147-8389
DOI: 10.1111/pace.14401
Popis: A retrospective analysis of 60 patients with hybrid ICD systems: Boston Scientific device paired with non-Boston leads. In 10 (17%) patients transient, out-of-range peaks of ventricular pace impedance trend were observed. Probable cause is header-lead interaction incompatibility. This matter is known mainly for pacemakers systems but not for ICDs. Investigation this issue is crucial because consequences in ICD systems are unpredictable and risk might be higher than in pacing systems. This article is protected by copyright. All rights reserved.