Popis: |
An analytical method of ultra-high performance liquid chromatography coupled to tandem mass spectrometry detection was developed and validated for the simultaneous quantification in plasma of four selective serotonin reuptake inhibitor antidepressants: sertraline, escitalopram, paroxetine, fluoxetine, and its metabolite norfluoxetine. A simple protein precipitation was performed with acetonitrile containing 100 ng/mL of indomethacin, which was used as internal standard. Chromatographic separation was carried out on an Acquity BEH C18 column with isocratic elution of the mobile phase consisting of 5 mmol/L ammonium acetate with 0.1% formic acid (A) and acetonitrile (B) at a 60:40 proportion, respectively. The flow rate was 0.4 mL/min with a run time of 5 min. A positive electrospray ionization source was used for detection. The method was linear in a range of 5-800 ng/mL, with determination coefficients greater than 0.991. The accuracy ranged from 91% to 112% for intra-assay and from 89% to 112% for inter-assay. The variation coefficients ranged from 3.1% to 14.88% for intra-assay and from 3.60% to 14.74% for inter-assay precision. The method was successfully applied for the analysis of 73 samples from patients under treatment with these antidepressants; 36.9% of the samples had concentrations outside therapeutic ranges. This method can be applied for routine analysis in clinical practice, simplifying sample processing, reducing analysis time and consequently the costs associated with it. |