Investigation of performance enhancement of flame photometric signal using a cylindrical impactor in the spray chamber

Autor: Ladislau Kékedy-Nagy
Rok vydání: 1997
Předmět:
Zdroj: Talanta. 44:1919-1922
ISSN: 0039-9140
DOI: 10.1016/s0039-9140(97)00074-x
Popis: The effect of a cylindrical impactor in the spray chamber of a flame photometer on the characteristics of the aerosol formed, on the transport-efficiency of the nebulizer-spray chamber system, as well as on the analytical signal has been investigated. The variables were: the diameter Phi of the cylindrical impactor and its position, i.e., the distance d in front of the tip of pneumatic nebulizer capillary where the impactor was fixed. The performance characteristics of an 1 mm Phi impactor was fixed at d=5 mm was superior to that of a conventional 8 mm Phi spherical bead.