Construction and properties of sub-miniature pressure pickups
Autor: | J. Van Der Ster, C. J. D. M. Verhagen, J. P. palm, W. Van Genderen, B. P. T. Veltman |
---|---|
Rok vydání: | 1954 |
Předmět: |
Materials science
Acoustics and Ultrasonics medicine.medical_treatment Acoustics General Chemical Engineering Linear variable differential transformer General Engineering General Physics and Astronomy Zero drift Atomic and Molecular Physics and Optics law.invention Core (optical fiber) Pressure measurement law medicine Physical and Theoretical Chemistry Instrumentation Mathematical Physics Cardiac catheterization |
Zdroj: | Applied Scientific Research. 3:409-416 |
ISSN: | 1573-1987 0003-6994 |
Popis: | Two manometers for dynamical measurements (up to frequencies of the order of 50 Hz) are described which were developed for measuring blood pressure in cardiac catheterization studies and for measurements in pneumatic controllers respectively. The principle of a differential transformer was used in which the core was entirely borne by a phosphorbronze membrane. The diameter of the blood pressure manometer is 2 mm (fitting to no. 6 catheter); its length 10 mm. The most important properties of the manometers are to be found in § 4. |
Databáze: | OpenAIRE |