Optimization of the exposure parameters in digital mammography for diverse glandularities using the contrast-detail metric

Autor: Oscar A. Martí Villarreal, Fermin G. Velasco, Agnes M.F. Fausto, Felix Mas Milian, Anderson W. Mol, Krizia R. Capizzi, Paulo Ambrosio
Rok vydání: 2022
Předmět:
Zdroj: Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 101
ISSN: 1724-191X
Popis: This investigation aims to study the optimization in digital mammography, considering a diverse percentage of breast glandularity using the contrast-detail metric.The Figure of Merit (FOM), defined as the ratio of the square of the Inverted Image Quality Figure (IQFIt was obtained new acquisition parameters for all glandularities that achieved a decrease in the MGD up to ∼ 50 %, maintaining the same image quality. The study was validated using the CIRS, TORMAM, and ACR phantoms through the contrast-to-noise ratio (CNR), the signal-to-noise ratio (SNR), and the MGD values obtained with the optimized parameters and the four AEC modes, which are the optimization proposed by the manufacturer.In this work, a new procedure was proposed that estimated the IQF
Databáze: OpenAIRE