Unsatisfying results in long-term atrial pacing with a bipolar active fixation atrial lead
Autor: | E. M. Buys, Johannes C. Kelder, Norbert M. van Hemel, Pascal F.H.M. van Dessel, Lex Bakema |
---|---|
Rok vydání: | 1994 |
Předmět: |
Male
medicine.medical_specialty Pacemaker Artificial Internal medicine Medicine Humans Heart Atria Lead (electronics) Bipolar lead Sick Sinus Syndrome Atrial pacing business.industry Follow up studies General Medicine Equipment Design Middle Aged Atrial Lead Surgery Heart Block Cardiology Female Cardiology and Cardiovascular Medicine business Heart atrium Active fixation Follow-Up Studies |
Zdroj: | Pacing and clinical electrophysiology : PACE. 17(12 Pt 1) |
ISSN: | 0147-8389 |
Popis: | A high dislodgment rate during long-term atrial pacing using the unipolar sickle-shaped active fixation lead was recently reported; therefore, the long-term results of atrial pacing in 118 consecutive patients with the bipolar sickle-shaped active fixation lead (Biotronik FH60-BP) were evaluated. Between January 1989 and September 1993, 87 leads (74%) were inserted for dual chamber pacing and 31 leads (26%) for atrial pacing only. At the time of implantation, the bipolar atrial electrogram had a mean voltage of 4.4 +/- 1.6 mV, whereas the acute atrial threshold was 0.72 +/- 0.38 V and 1.46 +/- 0.67 mA at 0.5-msec pulse duration and mean resistance 506 +/- 79 omega. Early lead dislodgment (1 month after implantation) occurred in 9 patients (7.6%). During a mean follow-up of 21.8 months (median 20.9 months), late dislodgment (1 month after implantation) occurred in 6 patients (5.1%) after a mean interval of 7.9 months (range 3-14 months). Due to the unacceptably high late dislodgment rate, which to date remains unexplained, new implants of this lead are not recommended. |
Databáze: | OpenAIRE |
Externí odkaz: |