Determination of bromate ion in drinking water by capillary zone electrophoresis with direct photometric detection

Autor: Shoji Motomizu, Makoto Ishida, Toshio Takayanagi, Rim Driouich, Julius Mbuna
Rok vydání: 2006
Předmět:
Zdroj: Journal of chromatography. A. 1128(1-2)
ISSN: 0021-9673
Popis: Bromate ion in drinking water was determined by capillary zone electrophoresis (CZE) with direct photometric detection. Bromate ion in the sample solution was introduced and concentrated into the capillary by electrokenetic injection for 50 s at −10 kV. Electrophoretic separation was made at an applied voltage of −25 kV and bromate ion was detected at wavelength 193 nm, at which the baseline was stabilized with less UV-absorbing acidic phosphate buffer. Bromate ion was detected within 5 min in the electropherogram. By increasing the electric conductivity in the migrating solution with 10 mM Na2SO4, a limit of detection (LOD) of 9 × 10−10 M (0.1 μg/L BrO3−) was achieved. The proposed method was applied to the analysis of tap water and river water samples, but bromate ion was not detected. Because the practical samples contain relatively large amount of foreign ionic substances, the tap water sample was diluted to avoid the matrix ions. Bromate ion added in a tap water at the concentration of 8 × 10−8 M was quantitatively recovered by diluting it 1/10.
Databáze: OpenAIRE