Quality of poly(somno)graphy recordings in children

Autor: Alessandro Amaddeo, Brigitte Fauroux, Lucie Griffon, Lynda Sidhoum, Sonia Khirani, Jorge Olmo Arroyo, Livio De Sanctis
Rok vydání: 2021
Předmět:
Zdroj: Journal of Sleep Research. 30
ISSN: 1365-2869
0962-1105
DOI: 10.1111/jsr.13241
Popis: The aim of the study was to assess the scorability of the signals of four poly(somno)graphy devices and transcutaneous carbon dioxide tracings (PtcCO2 ) of one device in children. The presence (0%, 2 years, respectively; nine poly(somno)graphies failed. For the four poly(somno)graphy devices, the signal was present during 99% of recording time for the electroencephalogram, 99% for thoracic belt, 97% for abdominal belt, 97% for body position, 95% for the microphone, 92% for pulse oximetry, 87% for tracheal sound, 71% for oronasal thermistor, 52% (41% for ≤ 2 years, 55% for > 2 years old) for nasal pressure and 86% for PtcCO2 . The signal was of good quality in 98% of poly(somno)graphies for body position, 96% for microphone, 96% for thoracic belt, 95% for pulse oximetry, 91% for abdominal belt, 91% for tracheal sound, 82% for oronasal thermistor, 78% for electroencephalogram, 73% for nasal pressure and 46% of PtcCO2 recordings. The scorability was comparable between devices. Nasal pressure and oronasal thermistor had the lowest scorability, especially in children aged ≤ 2 years. This underlines the necessity of the development or improvement of alternative, ideally face-free, sensors, or miniaturized devices adapted for infants and children.
Databáze: OpenAIRE