Evaluation of the Atellica COAG 360 coagulation analyzer in a central laboratory of a maximum care hospital

Autor: Andreas Peter, Sebastian Hörber, Rainer Lehmann
Rok vydání: 2019
Předmět:
Zdroj: Int. J. Lab. Hem., DOI: 10.1111/ijlh.13119 (2020)
ISSN: 1751-553X
1751-5521
DOI: 10.1111/ijlh.13119
Popis: Introduction: Fully-automated coagulation analyzers are key components of a high-throughput central laboratory. The novel Atellica COAG 360 (Siemens Healthineers) is a high-volume coagulation analyzer approved for hemostasis diagnostics. The aim of the study was to evaluate the analytical performance of this coagulation analyzer in a central laboratory. Methods: Intra (n=10)- and inter (n=20)-assay precision of the Atellica COAG 360 was determined using commercially available control samples. Patient samples (n=74-104) were used for comparison analyses with the Sysmex CS-5100 (Siemens Healthineers). Effects of visual interferences on coagulation testing were assessed and the sample throughput rate of the Atellica COAG 360 was determined. Results: Intra- and inter-assay precision of the Atellica COAG 360 showed coefficient of variations (CVs) 