Popis: We report a comparison of the results obtained at 500 MHz for heteronuclear shift correlation (HSQC) experiments with very small natural product samples using conventional and cryogenically cooled 3 mm NMR probes. The cryo probe affords a 12- to 16-fold reduction in data acquisition time for a comparable signal-to-noise ratio.