Comparison of Four Leukocyte Differential Methods with the National Committee for Clinical Laboratory Standards (NCCLS) Reference Method
Autor: | Linda M. Melchert, Bruce A. Hyma, Robert V. Pierre, Barbara A. Payne, Regina M. Scheidt, Won K. Lee |
---|---|
Rok vydání: | 1987 |
Předmět: |
Adult
Inflammation Pathology medicine.medical_specialty Leukemia Myeloproliferative Disorders Adolescent business.industry Anemia General Medicine Middle Aged Reference Standards Leukocyte Count Evaluation Studies as Topic Reference Values Reference values Leukocytes Screening method Humans Medicine Medical physics business Reference standards Aged Differential method |
Zdroj: | American Journal of Clinical Pathology. 87:201-209 |
ISSN: | 1943-7722 0002-9173 |
DOI: | 10.1093/ajcp/87.2.201 |
Popis: | The authors compared four leukocyte differential counting methods with the National Committee for Clinical Laboratory Standards Reference Leukocyte Differential Method H20-T to determine the clinical sensitivity of the methods. The three-part differential performed by the Coulter Counter Model S-Plus IV and the Toa E-5000, when combined with instrument flags and defined laboratory checking limits for red blood cell and platelet values, are safe and efficacious screening methods for the presence of morphologic abnormalities. The Geometric Data Hematrak 590 proved comparable in clinical sensitivity to a random 100-cell eye-count differential. |
Databáze: | OpenAIRE |
Externí odkaz: |