Monitoring the respiratory rate by miniature motion sensors in premature infants: a comparative study
Autor: | Anna Faingersh, S Kohn, Carmit Levy, Jimy Pesin, I C Klotzman, Avi Rotschild, Dan Waisman, G Hirshberg, Amir Landesberg |
---|---|
Rok vydání: | 2015 |
Předmět: |
Male
medicine.medical_specialty Pediatrics Respiratory rate High-Frequency Ventilation 03 medical and health sciences 0302 clinical medicine Respiratory Rate Internal medicine Diagnostic equipment medicine Electric Impedance Humans 030212 general & internal medicine Lung ventilation Motion sensors Diagnostic Equipment Monitoring Physiologic Respiratory Distress Syndrome Newborn Oscillatory ventilation business.industry Electrodiagnosis Infant Newborn Obstetrics and Gynecology Reproducibility of Results 030208 emergency & critical care medicine Equipment Design Respiratory support Dimensional Measurement Accuracy Visual Count Pediatrics Perinatology and Child Health Breathing Cardiology Female business Infant Premature |
Zdroj: | Journal of perinatology : official journal of the California Perinatal Association. 36(2) |
ISSN: | 1476-5543 |
Popis: | Existing respiratory rate (RR) monitors suffer from inaccuracy. The study assesses the accuracy of a novel modality that monitors lung ventilation with miniature motion sensors. RR was measured by three methods: impedance technology, motion sensors and visual count, in babies (n=9) that breathed spontaneously or with respiratory support and babies (n=12) that received high-frequency oscillatory ventilation (HFOV). A line close to equality (slope=0.96, r2=0.83) was obtained between the motion sensor and the visual count of the RR with narrow 95% limits of agreements ( |