Popis: The Dade Behring N Latex Cystatin C assay, a particle-enhanced nephelometric immunoassay for measuring serum cystatin C, was evaluated on the Dade Behring Nephelometer II. The assay time was 6 min and the throughput was 75 samples per hour. The sample volume was 40 microL and the measuring range was 0.25-7.90 mg/L. Imprecision studies revealed within-run CVs < 1.8% and between-run CVs < 1.8% in the concentration range 0.87-4.63 mg/L. Recovery was 92.4-101.3%. Linearity studies showed excellent correlation between the theoretical and obtained values. No interferences were detected from haemoglobin < 1.0 mmol/ L, bilirubin