Evaluation of the Dade Behring N Latex Cystatin C assay on the Dade Behring Nephelometer II System
Autor: | Else Randers, J. H. Kristensen, Erland J. Erlandsen |
---|---|
Jazyk: | angličtina |
Rok vydání: | 1999 |
Předmět: |
Clinical Biochemistry
Renal function Cysteine Proteinase Inhibitors urologic and male genital diseases Sample volume Serum cystatin Nephelometry and Turbidimetry medicine Humans Cystatin C Immunoassay Chromatography medicine.diagnostic_test biology Nephelometer Chemistry General Medicine Cystatins Calibration biology.protein Regression Analysis Cystatin Nephelometry |
Zdroj: | Erlandsen, E J, Randers, E & Kristensen, J H 1999, ' Evaluation of the Dade Behring N Latex Cystatin C assay on the Dade Behring Nephelometer II System ' Scandinavian Journal of Clinical and Laboratory Investigation. Supplement, vol. 59, no. 1, pp. 1-8 . Europe PubMed Central |
Popis: | The Dade Behring N Latex Cystatin C assay, a particle-enhanced nephelometric immunoassay for measuring serum cystatin C, was evaluated on the Dade Behring Nephelometer II. The assay time was 6 min and the throughput was 75 samples per hour. The sample volume was 40 microL and the measuring range was 0.25-7.90 mg/L. Imprecision studies revealed within-run CVs < 1.8% and between-run CVs < 1.8% in the concentration range 0.87-4.63 mg/L. Recovery was 92.4-101.3%. Linearity studies showed excellent correlation between the theoretical and obtained values. No interferences were detected from haemoglobin < 1.0 mmol/ L, bilirubin |
Databáze: | OpenAIRE |