Measuring Corneal Thickness with SOCT, the Scheimpflug System, and Ultrasound Pachymetry

Autor: Bartlomiej J. Kaluzny, Beata Soldanska, Mateusz Burduk, Józef Kałuzny, Ilona Piotrowiak
Jazyk: angličtina
Rok vydání: 2012
Předmět:
Zdroj: ISRN Ophthalmology
DOI: 10.5402/2012/869319
Popis: Background and Objective. Evaluation of agreement, repeatability, and reproducibility of central and minimal corneal thickness (CCT and MCT) measurements obtained by SOCT, the Scheimpflug system, and ultrasound pachymetry. Materials and Methods. 28 eyes of healthy patients were enrolled. Pachymetry measurements were performed with SOCT, the Scheimpflug system, and ultrasound instrument. Each measurement was taken by 3 operators on 3 devices providing a total of 2100 measurements. Results. The mean CCT for SOCT, Scheimpflug system, and ultrasound instrument was 537.92, 545.94, and 555.74 μm, respectively, (). The respective mean coefficients of repeatability for CCT were 0.61, 0.82 and 0.80, whereas mean coefficients of interoperator reproducibility for CCT were 0.91, 1.11, and 1.25. Conclusions. CCT and MCT measurements show moderate agreement between instruments. The repeatability and interoperator reproducibility of the results obtained by SOCT are somewhat higher. The operator's impact on CCT and MCT measurements is insignificant in all devices.