Radiation Dose to the Eye Lens through Radiological Imaging Procedures at the Surgical Workplace during Trauma Surgery
Autor: | Klaus Seide, Nils Weinrich, Christian Jürgens, Jens Dischinger, Heiko Martens, Christian Apelmann, Birgitt Kowald, Albert Nienhaus |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2019 |
Předmět: |
medicine.medical_specialty
Health Toxicology and Mutagenesis eye lens dose lcsh:Medicine Radiation Dosage Article 030218 nuclear medicine & medical imaging 03 medical and health sciences 0302 clinical medicine Occupational Exposure Lens Crystalline Humans Dosimetry Medicine Prospective Studies surgical assistant Radiometry Workplace trauma surgeon Surgeons Dosimeter business.industry lcsh:R Public Health Environmental and Occupational Health Hand surgery c-arm x-ray Radiation Exposure Dose area product fracture Surgical Procedures Operative 030220 oncology & carcinogenesis Radiological weapon Orthopedic surgery Wounds and Injuries Radiology sense organs hand surgeon Radiation protection dosis area product business eye dosimetry Trauma surgery |
Zdroj: | International Journal of Environmental Research and Public Health, Vol 16, Iss 20, p 3850 (2019) International Journal of Environmental Research and Public Health Volume 16 Issue 20 |
ISSN: | 1660-4601 |
Popis: | Background: Due to the drastic reduction of the eye lens dose limit from 150 mSv per year to 20 mSv per year since 2018, the prospective investigation of the estimated dose of the eye lens by radiological imaging procedures at the surgical site during trauma surgery in the daily work process was carried out. This was also necessary because, as experience shows, with changes in surgical techniques, there are also changes in the use of radiological procedures, and thus an up-to-date inventory can provide valuable information for the assessment of occupationally induced radiation exposure of surgical personnel under the current conditions. Methods: The eye lens radiation exposure was measured over three months for five trauma surgeons, four hand surgeons and four surgical assistants with personalized LPS-TLD-TD 07 partial body dosimeters Hp (0.07). A reference dosimeter was deposited at the surgery changing room. The dosimeters were sent to the LPS (Landesanstalt fü r Personendosimetrie und Strahlenschutzausbildung) measuring institute (National Institute for Personal Dosimetry and Radiation Protection Training, Berlin) for evaluation after 3 months. The duration of the operation, occupation (assistant, surgeon, etc.), type of surgery (procedure, diagnosis), designation of the X-ray unit, total duration of radiation exposure per operation and dose area product per operation were recorded. Results: Both the evaluation of the dosimeters by the trauma surgeons and the evaluation of the dosimeters by the hand surgeons and the surgical assistants revealed no significant radiation exposure of the eye lens in comparison to the respective measured reference dosimeters. Conclusions: Despite the drastic reduction of the eye lens dose limit from 150 mSv per year to 20 mSv per year, the limit for orthopedic, trauma and hand surgery operations is well below the limit in this setting. |
Databáze: | OpenAIRE |
Externí odkaz: |