medicine.medical_specialty Percutaneous business.industry medicine.medical_treatment Urology Lumen (anatomy) Surgery Urodynamics Ureter medicine.anatomical_structure Needles Nephrostomy medicine Pressure Humans Radiology Nuclear Medicine and imaging In patient Intrarenal pressure business Perfusion Nephrostomy Percutaneous Ureteral Obstruction |
Zdroj: | Radiology. 172(2) |
ISSN: | 0033-8419 |
Popis: | Ureteral perfusion studies in patients without preexisting renal access currently must be intermittently interrupted for intrarenal pressure measurement. A double-lumen needle has been successfully placed in four patients (two with native and two with transplanted kidneys). This permits simultaneous perfusion and intrarenal pressure monitoring yet maintains the safety and ease of use of a single skinny needle. |