The Accuracy of Continuous Glucose Monitoring in the Medical Intensive Care Unit
Autor: | Wannita Tingsarat, Nitchakarn Laichuthai, Patchaya Boonchaya-anant, Patinut Buranasupkajorn, Weerapan Khovidhunkit |
---|---|
Rok vydání: | 2021 |
Předmět: |
Blood Glucose
medicine.medical_specialty Continuous glucose monitoring business.industry Blood Glucose Self-Monitoring Critical Illness Endocrinology Diabetes and Metabolism Biomedical Engineering Reproducibility of Results Bioengineering Intensive Care Units Glucose Medical intensive care unit Technology Reports Emergency medicine Internal Medicine medicine Humans business Monitoring Physiologic |
Zdroj: | J Diabetes Sci Technol |
ISSN: | 1932-2968 |
DOI: | 10.1177/19322968211027590 |
Popis: | Objective: To assess the accuracy of continuous glucose monitoring (CGM) in medical intensive care unit (MICU) patients. Methods: A Medtronic Enlite® sensor accuracy was assessed versus capillary blood glucose (CBG) and plasma glucose (PG) using the mean absolute relative difference (MARD), surveillance error grid (SEG) analysis and modified Bland-Altman plots. Results: Using CBG as a reference, MARD was 6.6%. Overall, 99.7% of the CGM readings were within the “no risk” zone. No significant differences in accuracy were seen within vasopressor subgroups. Using PG as the reference, MARD was 8.8%. The surveillance error grid analysis showed 95.2% of glucose readings were within the “no risk” zone. There were no device-related adverse events. Conclusion: The CGM sensor showed acceptable accuracy in MICU patients, regardless of vasopressor use. |