Airway pressure triggered ventilation for preterm neonates

Autor: M. F. Hird, Anne Greenough, V. Chan
Rok vydání: 1991
Předmět:
Zdroj: jpme. 19:471-476
ISSN: 1619-3997
0300-5577
Popis: The usefulness of airway pressure triggered ventilation for the preterm newborn has been assessed using a new patient triggered valveless ventilator, the SLE 2000 infant ventilator (SLE 2000). This ventilator performs well at fast rates with no inadvertent positive end expiratory pressure (PEEP) even at rates of 150 breaths per minute (bpm). The ventilator is triggered by a change in airway pressure equal to or exceeding 0.5 cmH2O. If the infant fails to achieve the change in airway pressure which will trigger the ventilator the infant is ventilated at the back-up rate which is predetermined in conventional mode prior to commencing PTV. Infants were ventilated for one hour on a conventional neonatal ventilator, then for one hour on the SLE 2000 in conventional mode without changing the ventilator settings and finally for one hour on the SLE 2000 in patient triggered mode. Arterial blood gases were checked at the end of each hour. During patient triggered ventilation (PTV) the peak pressure, inspiratory time and inspired oxygen concentration were the same as those used during conventional mode. Simultaneous recordings were made of flow, volume, ventilator and oesophageal pressure change, from this recording the trigger delay during PTV was calculated. The trigger delay, being the time lag from the start of spontaneous inspiration, indicated by the negative deflection in the oesophageal pressure trace, and the onset of the ventilator breath. Thirteen infants were included in the study, median gestational age 32 weeks (range 25-35) and birthweight 1640 g (range 838-3038). All were being ventilated for respiratory distress syndrome (RDS) and were 4 days of age.(ABSTRACT TRUNCATED AT 250 WORDS)
Databáze: OpenAIRE