Long-term reliability of single lead atrial synchronous pacing systems using closely spaced atrial dipoles: five-year experience
Autor: | Gian Franco Percoco, Gabriele Guardigli, Gian Enrico Antonioli, L. Ansani, Tiziano Toselli |
---|---|
Rok vydání: | 1994 |
Předmět: |
Male
medicine.medical_specialty Pacemaker Artificial Single Center Electrocardiography Pacemaker replacement Internal medicine medicine Humans Pacemaker malfunction Lead (electronics) Aged Aged 80 and over business.industry General Medicine Ventricular pacing Middle Aged Heart Block Single lead Cardiology Vdd pacing Equipment Failure Female Implant Cardiology and Cardiovascular Medicine business Follow-Up Studies |
Zdroj: | Pacing and clinical electrophysiology : PACE. 17(11 Pt 2) |
ISSN: | 0147-8389 |
Popis: | To assess the long-term capability of single atrioven ticular (AV) lead VDD pacing systems using close atrial dipoles to assure reliable atrial guided pacing, the safety and efficacy of 86 VDD units implanted in 73 patients at a single center since November 1988 was reviewed. All patients suffered from advanced AV block with normal sinoatrial function. Sixty five patients received a LEM/CCS Twinal 30/30S system, four patients received a Vitatron-Saphir system, and four patients received a Medtronic Thera VDR 8348 system. All patients underwent provocative tests in search of myopotential interference, and Holter recordings; in a group of patients who underwent pacemaker replacement a comparison was made between implant and replacement measurements. The mean follow-up duration was 27.3 months. A high percentage of successfully VDD paced patients and a low incidence of pacemaker malfunction, regularly solved by pacemaker reprogramming, was reported. Atrial signal amplitudes comparable to those measured at implant were found at replacement in all patients. These data support the long-term reliability of single AV lead VDD pacing systems with closely spaced atrial dipoles, as well as stable atrial sensing by floating bipolar atrial electrodes and effective atrial synchronous ventricular pacing over time. |
Databáze: | OpenAIRE |
Externí odkaz: |