Intermittent output failure of a VVI device due to the disintegration of the generator

Autor: Arnold Pintér, István Préda, Sándor Kayser, Ádám Böhm
Rok vydání: 2001
Předmět:
Zdroj: Pacing and clinical electrophysiology : PACE. 24(1)
ISSN: 0147-8389
Popis: BOHM, A., et al.: Intermittent Output Failure of a VVI Device Due to the Disintegration of the Generator. The metal case and the header of a Biovallees Coralite 247 pacemaker fell apart in a patient resulting in syncope due to myopotential inhibition. Upon recalling our patients with that type of device, we found a similar technological failure in 11 of 16 of them. This finding warrants a consideration for recalling that Biovallees device.