Intermittent output failure of a VVI device due to the disintegration of the generator
Autor: | Arnold Pintér, István Préda, Sándor Kayser, Ádám Böhm |
---|---|
Rok vydání: | 2001 |
Předmět: |
Male
medicine.medical_specialty Pacemaker Artificial Generator (computer programming) business.industry General Medicine Syncope Equipment Failure Analysis Electric Power Supplies Internal medicine Header medicine Cardiology Electrocardiography Ambulatory Product Surveillance Postmarketing Humans Equipment Failure Cardiology and Cardiovascular Medicine business Syncope (phonology) Aged |
Zdroj: | Pacing and clinical electrophysiology : PACE. 24(1) |
ISSN: | 0147-8389 |
Popis: | BOHM, A., et al.: Intermittent Output Failure of a VVI Device Due to the Disintegration of the Generator. The metal case and the header of a Biovallees Coralite 247 pacemaker fell apart in a patient resulting in syncope due to myopotential inhibition. Upon recalling our patients with that type of device, we found a similar technological failure in 11 of 16 of them. This finding warrants a consideration for recalling that Biovallees device. |
Databáze: | OpenAIRE |
Externí odkaz: |