Effect of Dose Reduction on the Ability of Digital Mammography to Detect Simulated Microcalcifications

Autor: Yoshio Matsuo, Mayumi Cho, Mari Yakabe, Masafumi Masuda, Hidetake Yabuuchi, Taro Setoguchi, Takeshi Kamitani, Masayuki Sasaki, Shuji Sakai
Jazyk: angličtina
Rok vydání: 2009
Předmět:
Popis: The purpose of this article was to report the relationship between radiation dose and the ability of sentence digital mammography to detect microcalcifications. All images were acquired by computed radiography and an anthropomorphic breast phantom. The tube voltage and anode/filter combination used were 28 kVp and Mo/Mo. Simulated microcalcifications with an approximate diameter of 250–350 μm were positioned on the phantom. Groups of six microcalcifications were arranged in one of two patterns, a line cluster 1 cm long or a hexagonal cluster 4 mm wide. One of the six microcalcifications was removed to create a negative control. Each cluster was placed on 25 different points. Four levels of milliampere-second (mAs) values were applied: 100%, 50%, 25%, and 12.5%. Five staff radiologists participated in an observer performance test. All observers used a workstation with a 3-megapixel monochrome LCD monitor. The areas under the receiver-operating characteristics curves (AUC) were used to compare diagnostic performance among the four doses. The overall AUC scores were 0.97 with 100% mAs, 0.93 (n.s.) with 50%, 0.90 (p