Survey of patient dose in computed tomography in Syria 2009
Autor: | Mohammad Hassan Kharita, S. Khazzam |
---|---|
Rok vydání: | 2010 |
Předmět: |
Adult
medicine.medical_specialty Computed tomography Radiation Dosage Ct dose index Risk Assessment Effective dose (radiation) Pelvis Reference Values Reference level medicine Humans Radiology Nuclear Medicine and imaging Medical physics Child Radiometry Radiation Syria Radiological and Ultrasound Technology medicine.diagnostic_test Public Health Environmental and Occupational Health International Agencies General Medicine Collective dose Health Surveys Geography Reference values Ct scanners Patient dose Tomography X-Ray Computed Head Relative Biological Effectiveness |
Zdroj: | Radiation Protection Dosimetry. 141:149-161 |
ISSN: | 1742-3406 0144-8420 |
DOI: | 10.1093/rpd/ncq155 |
Popis: | The radiation doses to patient in computed tomography (CT) in Syria have been investigated and compared with similar studies in different countries. This work surveyed 30 CT scanners from six different manufacturers distributed all over Syria. Some of the results in this paper were part of a project launched by the International Atomic Energy Agency in different regions of the world covering Asia, Africa and Eastern Europe. The dose quantities covered are CT dose index (CTDI(w)), dose-length product (DLP), effective dose (E) and collective dose. It was found that most CTDI(w) and DLP values were similar to the European reference levels and in line with the results of similar surveys in the world. The results were in good agreement with the UNSCEAR Report 2007. This study concluded a recommendation for national diagnostic reference level for the most common CT protocols in Syria. The results can be used as a base for future optimisation studies in the country. |
Databáze: | OpenAIRE |
Externí odkaz: |