Construction of a fine wire thermocouple capable of repeated insertions into and accurate positioning within a controlled environmental chamber

Autor: N. Todreas, L. Battist, F. Goldner
Rok vydání: 1969
Předmět:
Zdroj: Medical & Biological Engineering. 7:445-448
ISSN: 1741-0444
0025-696X
Popis: A method of construction is presented for fabrication and incorporation of a fine wire exposed junction thermocouple into a hypodermic syringe. The junction is protected by the hypodermic needle until the needle has been placed in a desired location. The 0·0005 in. chromal-constantan thermocouple formed has a bead diameter of approximately 30 μ (3×10−3 cm), a time constant in still water of less than 1 msec, and one of the largest thermoelectric powers (V/degree) of the commercially available thermocouple wire combinations.