Measuring Vibration Parameters for Ultrasonic Contact Lithotripters
Autor: | A. M. Enyakov |
---|---|
Rok vydání: | 2016 |
Předmět: |
Test bench
Vibration acceleration Materials science Hydrophone business.industry Applied Mathematics Acoustics Ultrasound 030232 urology & nephrology Vibration velocity Vibration 03 medical and health sciences 0302 clinical medicine 030220 oncology & carcinogenesis Physics::Atomic and Molecular Clusters Physics::Accelerator Physics Waveguide (acoustics) Ultrasonic sensor Physics::Chemical Physics business Instrumentation |
Zdroj: | Measurement Techniques. 59:303-309 |
ISSN: | 1573-8906 0543-1972 |
Popis: | Methods are considered for measuring the parameters of ultrasonic contact lithotripters with waveguide ends executing low-amplitude longitudinal vibrations. A method is proposed for measuring vibration acceleration, vibration velocity, and the range of these vibrations using a hydrophone calibrated on a vibration test bench. |