Effect of slice thickness on image noise and diagnostic content of single-source-dual energy computed tomography
Autor: | Marwan Alshipli, Norlaili A. Kabir |
---|---|
Rok vydání: | 2017 |
Předmět: |
History
medicine.medical_specialty Computer science business.industry Visibility (geometry) Digital Enhanced Cordless Telecommunications Dual-Energy Computed Tomography Imaging phantom 030218 nuclear medicine & medical imaging Computer Science Applications Education 03 medical and health sciences Noise DICOM 0302 clinical medicine Software 030220 oncology & carcinogenesis Image noise medicine Radiology business Biomedical engineering |
Zdroj: | Journal of Physics: Conference Series. 851:012005 |
ISSN: | 1742-6596 1742-6588 |
DOI: | 10.1088/1742-6596/851/1/012005 |
Popis: | Computed tomography (CT) employs X-ray radiation to create cross-sectional images. Dual-energy CT acquisition includes the images acquired from an alternating voltage of X-ray tube: a low- and a high-peak kilovoltage. The main objective of this study is to determine the best slice thickness that reduces image noise with adequate diagnostic information using dual energy CT head protocol. The study used the ImageJ software and statistical analyses to aid the medical image analysis of dual-energy CT. In this study, ImageJ software and F-test were utilised as the combination methods to analyse DICOM CT images. They were used to investigate the effect of slice thickness on noise and visibility in dual-energy CT head protocol images. Catphan-600 phantom was scanned at different slice thickness values;.6, 1, 2, 3, 4, 5 and 6 mm, then quantitative analyses were carried out. The DECT operated in helical mode with another fixed scan parameter values. Based on F-test statistical analyses, image noise at 0.6, 1, and 2 mm were significantly different compared to the other images acquired at slice thickness of 3, 4, 5, and 6 mm. However, no significant differences of image noise were observed at 3, 4, 5, and 6 mm. As a result, better diagnostic image value, image visibility, and lower image noise in dual-energy CT head protocol was observed at a slice thickness of 3 mm. |
Databáze: | OpenAIRE |
Externí odkaz: |