A detector for intermediate energy neutrons

Autor: J.K. Basson
Rok vydání: 1963
Předmět:
Zdroj: Nuclear Instruments and Methods. 22:339-344
ISSN: 0029-554X
DOI: 10.1016/0029-554x(63)90261-1
Popis: An intermediate energy neutron detector has been developed, consisting of a small wax sphere surrounding a boron-loaded scintillation counter. The energy dependence of the detector was tested using the Harwell “neutron booster”, which provides short pulses of neutrons whose energies may be measured by time-of-flight techniques. With the optimum size of 6.3 cm diameter, the response was proportional to within ± 12% of the neutron dose rate in the energy range 1 eV to 10 keV. This size would be suitable for use in a portable survey monitor.