Evaluation and Comparison of Two Total Protein Reagents
Autor: | Scott K. Steinfeldt, Leonard Gary Nielsen |
---|---|
Rok vydání: | 1986 |
Předmět: | |
Zdroj: | Laboratory Medicine. 17:82-84 |
ISSN: | 1943-7730 0007-5027 |
DOI: | 10.1093/labmed/17.2.82 |
Popis: | Sigma Diagnostics total protein reagent was evaluated in the authors' laboratory as an alternative to Abbott A-Gent total protein reagent, for use on the ABA-100 chemistry analyzer. The Sigma reagent at least equalled the performance of the Abbott reagent for precision, accuracy, and recovery. A high degree of correlation (r = .942) was found between the results obtained using the two reagents. The Sigma reagent offered significantly longer stability and lower cost. |