Autor: |
Y. Murayama, A. Iwasa, T. Sakuraba, Y. Sakamoto, T. Endo, H. Yoshida, T. Kozakai |
Rok vydání: |
1997 |
Předmět: |
|
Zdroj: |
IEEE Transactions on Instrumentation and Measurement. 46:233-236 |
ISSN: |
0018-9456 |
DOI: |
10.1109/19.571820 |
Popis: |
A voltage standard system using a 10 V Josephson junction array has been developed. The uncertainty is 0.6/spl times/10/sup -8/ for measurement of the 10 V output of a Zener diode reference standard. A comparison of the 10 V array system with the conventional 10 V measurement system composed of a 1 V array and a voltage divider was performed through measurements of a 10 V output of a Zener. The result agreed within the uncertainty (2/spl times/10/sup -8/) of the comparison. The 10 V output voltages of a Zener reference standard are measured consecutively by the 10 V array system. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|