Relevant reduction effect with a modified thermoplastic mask of rotational error for glottic cancer in IMRT

Autor: Kwang Hwan Cho, Mi Ryeong Ryu, Jae Hong Jung, Yong Ho Kim, Tae Suk Suh, Joo-Young Jung, Sun Hyun Bae, Bo-Young Choe, Seong Kwon Moon
Rok vydání: 2017
Předmět:
Zdroj: Journal of the Korean Physical Society. 70:308-316
ISSN: 1976-8524
0374-4884
DOI: 10.3938/jkps.70.308
Popis: The purpose of this study was to analyze the glottis rotational error (GRE) by using a thermoplastic mask for patients with the glottic cancer undergoing intensity-modulated radiation therapy (IMRT). We selected 20 patients with glottic cancer who had received IMRT by using the tomotherapy. The image modalities with both kilovoltage computed tomography (planning kVCT) and megavoltage CT (daily MVCT) images were used for evaluating the error. Six anatomical landmarks in the image were defined to evaluate a correlation between the absolute GRE (°) and the length of contact with the underlying skin of the patient by the mask (mask, mm). We also statistically analyzed the results by using the Pearson’s correlation coefficient and a linear regression analysis (P