Autor: |
William N. Hogenauer, Sangho Kim, Kenneth Kensey, Young I. Cho |
Rok vydání: |
2002 |
Předmět: |
|
Zdroj: |
Journal of Non-Newtonian Fluid Mechanics. 103:205-219 |
ISSN: |
0377-0257 |
DOI: |
10.1016/s0377-0257(02)00004-6 |
Popis: |
The present study introduces a method to consider different surface tensions at two riser tubes for the viscosity measurements of fluids using a U-shaped scanning capillary-tube viscometer (SCTV). The accuracy and repeatability of the SCTV were demonstrated by comparing the viscosity results of distilled water obtained from the SCTV with the well-accepted reference data for water. Non-Newtonian viscosities of bovine blood containing 7.5% ethylenediaminetetraacetic acid (EDTA) and fresh human blood were measured at 25 °C and at a body temperature of 37 °C, respectively. The viscosity results for bovine blood gave excellent agreement with the results obtained from a rotating viscometer. The results demonstrate that whole blood viscosity can be measured over a range of shear-rates in less than 2–3 min without any anticoagulants using disposable U-shaped capillary-tube sets. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|