Popis: |
For the determination of trace matals in a sample, such as rain water, contamination from the supporting electrolyte solution and the standard solution are not negligible. In this study, the relationship between the composition of the solution in the statndard addition method and the response signal was analyzed. It was concluded that two correction terms for eliminating contamination were added to the usual formula used to determine. The trace metals, such as Cu(II), Pb(II), Cd(II) and Zn(II) in rain water were determined by anodic stripping voltammetry with the supporting eletrolyte solution containing 7.68×10-2 M tartaric acid and 3.84×10-2 M ammonium acetate. The proposed method can be used to determine ppb level of trace metals with good precision. |