HURWA Robotic-Assisted TKA System is Suitable for a Variety of Prosthesis Types

Autor: Xingye Li, Jun Yan, Xiaofeng Zhang, Shugang Li, Zheng Li, Lele Ding
Rok vydání: 2021
Předmět:
DOI: 10.21203/rs.3.rs-1086121/v1
Popis: Background Total knee arthroplasty (TKA) is an effective and also mature surgical interventions that improves life quality and provides pain relief. Accurate bone cuts are important to prevent TKA malalignment and it requires cautious preoperative plan and precise bone resection. Recently, robotic-assisted TKA techniques have been used to improve the accuracy of bone resection and implantation. However, the system described above suits for only one prosthesis type. Methods Five types (MicroPort_CS, Smith& Nephew_GII, Johnson&Johnson_PFC_PS, kingnow _VLQX_PS and Akmedical_A3GT_PS) implants were included in our study and three Sawbones models were used for each implant. Procedures were performed by experienced joint replacement surgeons using HURWA robotic-assisted TKA system. Results our study indicated that the bone resection error of HURWA robotic-assisted system was below 0.5 mm (with SDs below 0.3 mm), and all of the bone resection angles were below 0.5° (with SDs below 0.3°). The bone resection angles and levels deviation of different brand prosthesis types were below 0.5 mm (with SDs below 0.3 mm) and below 0.5° (with SDs below 0.3°) respectively. Conclusion It suggested that our system may be suitable for different prosthesis types.