Point projection radiography with the flexible x-ray imager

Autor: Ted Perry, D. Hargrove, A. Nikitin, J. R. Mazuch, S. A. Alvarez, K. S. Budil, Perry M. Bell
Rok vydání: 1997
Předmět:
Zdroj: Review of Scientific Instruments. 68:796-798
ISSN: 1089-7623
0034-6748
DOI: 10.1063/1.1147744
Popis: Radiography techniques utilizing large-area x-ray sources (typically ⩽7 keV) and pinhole-imaging gated x-ray diagnostics have long been used at the Nova laser facility. However, for targets requiring higher-energy x-ray backlighters (>9 keV), low conversion efficiencies and pinhole losses combine to make this scheme unworkable. The technique of point projection radiography has been improved upon to make imaging at high x-ray energies feasible. In this scheme a “point” source of x-rays, usually a small diameter (⩽25 μm) fiber, is illuminated with a single, 100-ps pulse from the Nova laser. A gated x-ray imager with a 500-ps electronic gate width is used to record the projected image. [K. S. Budil et al., Rev. Sci. Instrum. 67, 485 (1996).] The experimental challenges this technique presents and experimental results will be discussed.