Determination of trace amounts of chromium by atomic absorption spectrometry with a tantalum filament atomizer
Autor: | Tsugio Takeuchi, Toshihisa Maruta |
---|---|
Rok vydání: | 1973 |
Předmět: |
inorganic chemicals
Condensed Matter::Quantum Gases Accuracy and precision Argon Trace Amounts Physics::Instrumentation and Detectors technology industry and agriculture Tantalum Analytical chemistry chemistry.chemical_element Biochemistry Analytical Chemistry law.invention Protein filament Chromium chemistry law Physics::Atomic and Molecular Clusters Environmental Chemistry Physics::Atomic Physics Atomic absorption spectroscopy Absorption (electromagnetic radiation) Spectroscopy |
Zdroj: | Analytica Chimica Acta. 66:5-11 |
ISSN: | 0003-2670 |
DOI: | 10.1016/s0003-2670(01)80196-3 |
Popis: | The sample solution containing chromium was vaporized and atomized from a tantalum filament by electrical heating into an argon stream within an absorption chamber. The interference of various metals on the atomic absorption spectrometry of chromium was investigated. Interferences were minimized by using a high filament temperature. The flameless atomizer was used for the determination of chromium in steel at a filament temperature of 2150°; with the steels investigated no matrix effect was observed, and the accuracy and precision were satisfactory. |
Databáze: | OpenAIRE |
Externí odkaz: |