Electronic readout of a-Si EPIDs for optimum signal-to-noise ratio

Autor: Jean Pouliot, Farhad A. Ghelmansarai, Satrajit Misra
Rok vydání: 2003
Předmět:
Zdroj: SPIE Proceedings.
ISSN: 0277-786X
DOI: 10.1117/12.480481
Popis: Three acquisition schemes for a-Si flat panels are described for radiation therapy imaging. The goal of all three acquisition modes is to acquire images with the highest achievable SNR (signal to noise ratio). The acquisition modes are Single mode for low dose acquisition (used for patient positioning), external continuous mode used for patient treatment (verification), and Cone Beam mode for mega-voltage computed tomography (MVCT). During single mode acquisition, a few frames are readout prior to the start of irradiation. During this cycle, the accumulated dark current and residual data are cleared. During the radiation delivery no readout occurs, and the signal is integrated over the entire exposure period. After the irradiation readout occurs. The advantages of this readout scheme are to reduce the effects of readout noise and eliminate the linear accelerator (linac) pulsing effects on the final image. There is no readout during the exposure; therefore, no beam pulsing artifacts occur. Since the signal is integrated during the exposure time and the readout is performed after the exposure, this improves the SNR compared to acquiring a few frames during the radiation and averaging these frames to create the final image. The single mode acquisition is used clinically routinely and allows the acquisition of clinical images with a small amount of exposure (