Establishment of CZT pixellated radiation detection system and pulse-height correction
Autor: | Min Shen, Sha li Xiao, Miao Li |
---|---|
Rok vydání: | 2018 |
Předmět: |
010302 applied physics
Physics Pixel Physics::Instrumentation and Detectors 010308 nuclear & particles physics business.industry Resolution (electron density) Detector 01 natural sciences Atomic and Molecular Physics and Optics Particle detector Electronic Optical and Magnetic Materials Full width at half maximum Optics Pulse-amplitude modulation Computer Science::Computer Vision and Pattern Recognition Computer Science::Multimedia 0103 physical sciences Electrical and Electronic Engineering business Spectroscopy Energy (signal processing) |
Zdroj: | Optik. 158:1220-1226 |
ISSN: | 0030-4026 |
Popis: | 11 × 11 × 3 mm3 and 4 × 4 pixel array CdZnTe (CZT) detector system based on the principle of CZT pixel array detector and correction algorithm has been established. The leakage current of CZT pixellated crystal is tested and analyzed. The pulse amplitude spectrum of the CZT detector has been measured using 137 Cs 662 keV Gamma source. The average energy resolution of pixellated detector system is 3.3% and full energy peak FWHM is 21.85 keV. The corrected algorithm of pulse-height is applied for improving spectroscopy response. The average energy resolution of corner pixels has been improved. The best energy resolution of pixels is 2.5% after correction. The energy resolution of the 4 × 4 pixellated CZT detector system is excellent. The spectrum performance of edge pixels is obvious improved by pulse-height correction. The resolution of edge pixels is improved efficiently. |
Databáze: | OpenAIRE |
Externí odkaz: |